BACK OF THE METER

Figure 2-2 shows the connectors on the back of the meter. Table 2-2 gives a brief description of each connector at the back of the meter. Refer to Figure 2-3 for dc-power connection information.

Connectors (AC-powered)
Figure 3.1
Connectors (AC-powered)
Connectors (DC-powered)
Figure 3.2
Connectors (DC-powered)

Connector
Description
J1-1
J1-2
J1-3
AC high connection (NC on DC-powered units)
AC low connection (+ input on DC-powered units)
AC ground (DC-power return on DC-powered units).
J2-1


J2-2
J2-3
Display hold input (active low) --connect to J2-2 to hold meter reading. Peak and valley readings are continuously updated during display hold even though display is on hold.
Ground.
Display blank input (active low) -- connect to J2-2 to blank meter display.
J4-1
J4-2
+ TC input.
- TC input (red wire)

Main Board AC Power Jumpers
To check voltage jumpers, or to change from 115 V to 230 Vac:
  • Disconnect the AC power from the meter.

  • Remove the front lens, if present.

  • Pull the meter forward, out of its case.


  • 115Vac Jumpers
    Figure 3.3 115Vac Jumpers (Default)
    230Vac Jumpers
    Figure 3.4 230Vac Jumpers

    S51, -S4 Jumpers
    The S51 Jumpers is at the front of the meter near the digital isplay. The S4 Jumpers is on the side of the main board (see igure 3.5). (Defaults are in bold and italic.)
  • Disconnect the main power from the meter.

  • Remove the front lens, if present.

  • Pull the meter forward, out of its case.


  • S51-S4 Jumpers Table 3.2
    Jumper
    Description
    S51
    installed:
    removed:
    front panel buttons locked out
    all buttons operable. Jumper placed in storage position on one PIN only.
    S4
    installed:
    removed:
    calibration enabled
    calibration disabled. Jumper placed in storage position on one PIN only.

    S51, -S4 Jumpers
    Figure 3.5 S51, -S4 Jumpers

    General Specifications
    Microprocessor based
    Input type: Type J, DINJ, K, T, E, N, R, S
    Max Error: ▒ 1 deg C (▒ 1.8 ºF)
    Accuracy at 25 ºC: ▒ .5 deg C (▒ .8 ºC for DINJ)
    Cold-junction compensation tempco: 0.07ºC/ºC


    Thermocouple Types Range
    Range
    Type
    Deg C
    DegF
    J
    -210
    -346
    760
    1400
    DIN J
    -200
    -328
    900
    1652
    K
    -270
    -454
    1372
    2500
    T
    -270
    -454
    400
    752
    E
    -270
    -454
    1000
    1832
    R/S
    -50 -58
    1768 3214
    N
    -270
    -454
    1300
    2372
    Display
    Type: 7 segments red or green 4 digit
    Height: 0.56in (14.2mm)
    Resolution:
    0.1º or 1º resolution

    Power
  • 115 or 230 Vac
  • ▒15% (2 Watt max.)

  • Frequency
    50 or 60Hz

  • Isolation
  • 354V peak per IEC spacing
  • NMR 60dB
  • CMR 120 dB


  • DC voltage
    10-32Vdc isolated to 300Vp

    Environmental
    Operating temp: 0╨60ºC
    Storage temp: -40 to +85ºC
    Relative humidity: 95% at 40ºC (non condensing)


    Case
    Newport DIN 4896-100

    Panel cutout
    1.77 X 3.62 in (45 X 92mm)

    Depth behind bezel:

    3.94 in (100mm)

    Material
    94V-0 UL-rated polycarbonate

    Meter Dimensions Figure 3.6
    Dimensions
    Dimensions- Panel Thickness
    Help Price List Specifications Print Manual Manuals Home